Instrument Formats & QC

One page, three questions per instrument: what does the raw file look like, how is an instrument error decided, and where does that decision get recorded?

This is the reference for everything L2 does — the parsing recipe and the error/status semantics — for when a file won't parse, when Status Error fires on every row, or when it never fires at all. The per-instrument API pages cover usage; the level model that frames all of this is in Data Levels (L0–L3), whose rules R2 and R2a define what QC may and may not do to your data.


1. How a status judgement is made

Every status check goes through one function — QualityControl.filter_error_status (core/qc.py) — in one of four modes. The mode determines how the raw status value is interpreted and how the ignored_status_errors whitelist is applied.

Mode Raw value looks like Error when Instruments ignored_status_errors entries are…
bitwise integer, OR-summed bits (0, 4, 536870912) any non-whitelisted code in ERROR_STATES matches bitwise, or the value equals a special_codes entry AE33, AE43, BC1054, MA350, TEOM integer codes/bits removed from the error definition
numeric flat integer code (0, 4, 16) status != ok_value and not NaN Aurora, NEPH numeric codes treated as OK
text free text, possibly comma-joined (Normal Scan, Low aerosol flow,Neutralizer not active) any token is neither ok_value nor whitelisted SMPS (six columns across two AIM dialects) string tokens, matched per token
binary_string space-grouped bit string (0000 0000 0000 0000) any non-whitelisted bit remains set after clearing the whitelist mask APS integer bit masks cleared before testing

Two behaviours to know:

  • Empty means OK, always. In text mode the sentinels '', 'nan' and 'None' are never errors. ('None' arrives when a column is missing from some files of a multi-file concat: pandas fills with Python None, which astype(str) renders as 'None'.)
  • A missing status column reads as OK — but now says so. filter_error_status returns all-False for a column it cannot find, which is deliberate (mixed exports must not crash) but used to make a status rule completely inert with no signal. Every status reader now calls check_status_columns first and logs a warning naming what it looked for and what the file actually contains, so a new export dialect surfaces instead of silently passing every fault.

Entries in ignored_status_errors that don't fit a mode are skipped, so the same whitelist can be passed to any reader safely:

RawDataReader('TEOM', path, ignored_status_errors=[536870912])          # bitwise: ignore "Dryer A"
RawDataReader('SMPS', path, ignored_status_errors=['Low aerosol flow']) # text: per-token
RawDataReader('APS',  path, ignored_status_errors=[1, 2])               # binary_string: clear bits 0,1
RawDataReader('Aurora', path, ignored_status_errors=[4, 16])            # numeric: extra OK codes

2. Where a judgement is recorded

Channel Written by Granularity Content
QC_Flag column _QC / _process (L2) per row "Valid", or comma-joined names of every rule that fired (advisory ones included)
QC_Invalid column _QC / _process (L2) per row boolean verdict: did an invalidating rule fire? The only thing masking looks at
{inst}.log all levels per run parse warnings, skipped/dropped files, mixed-resolution & stray-date warnings, QC Summary (count + % per rule), rates
report.json L3 per week / month + timeline rates.weekly / rates.monthly (acquisition / yield / total), timeline of operational & down periods with reasons
df.attrs L3 (_stamp) per call provenance, coverage_* vs requested_*, raw_freq, freq_mixed, acquisition_rate / yield_rate / total_rate
_read_{inst}_qc.csv L2 per row the flagged frame before masking — the only place where flag and value coexist

Lifecycle of one flag:

_QC       → QCRule.condition(df) → mask → QC_Flag   = "Below MDL, Status Error"
                                        → QC_Invalid = True   (Status Error is severity='error';
                                                               Below MDL alone would leave it False)
_process  → may append via update_qc_flag(..., severity=...) → "…, Invalid AAE"
log       → "  Status Error: 24312 (4.9%)" / "  Below MDL: 1201 (2.4%) [advisory]"
report    → yield counts rows that survived the verdict, so an advisory-only row counts
__call__  → QC_Invalid  →  entire row set to NaN; QC_Flag + QC_Invalid dropped

Not every flag is fatal

Each rule carries a severity. 'error' (the default) masks the row; 'warning' records the flag and keeps the measurement. Advisory today: OCEC's Below MDL and Xact's Upscale Warning — a sub-detection-limit value or a vendor upscale notice is a real measurement, and masking is per row, so invalidating it would delete every other species at that timestamp too.

Insufficient is advisory too: it describes how well an hour is covered, which decides whether an average over it is representative, not whether the readings are real. Spike remains invalidating — it asserts the value itself is wrong. Reclassify per run without editing a reader:

# tighten an advisory flag back into an invalidating one
RawDataReader('SMPS', path, flag_severity={'Insufficient': 'error'})

To see values behind an invalidating flag, read _read_{inst}_qc.csv (flag, verdict and value side by side) or call with qc=False (returns L1 — no QC, and no resampling either).

QC_Flag vocabulary

Severity is error unless marked advisory.

Flag Meaning Used by
Valid passed every rule all
Status Error instrument status register reports a non-whitelisted condition AE33, AE43, BC1054, MA350, SMPS, APS, Aurora, NEPH, TEOM
Insufficient an hour holds < 50 % of the points it could have held, given how much of that hour the read covers — advisory AE33, AE43, BC1054, MA350, SMPS, APS, Aurora, NEPH, TEOM, GRIMM
Invalid BC any BC channel ≤ 0 or > 20 000 ng/m³ AE33, AE43, BC1054, MA350
Invalid AAE |AAE| outside 0.7–2.0 (added in _process) AE33, AE43, BC1054, MA350
Invalid Number Conc total number concentration outside range (SMPS 2 000–1e7, APS 1–700 #/cm³) SMPS, APS
DMA Water Ingress any bin > 400 nm exceeds 4 000 dN/dlogDp SMPS
Invalid Scat Value any scattering channel ≤ 0 or > 2 000 Mm⁻¹ Aurora, NEPH
Invalid Scat Rel B < G < R — inverted wavelength dependence Aurora, NEPH
No Data every measurement channel NaN Aurora, NEPH, GRIMM
High Noise TEOM noise ≥ 0.01 TEOM
Non-positive PM_NV ≤ 0 or PM_Total ≤ 0 TEOM
NV > Total PM_NV > PM_Total (physically impossible) TEOM
Spike change > 3 × median absolute change, or a large up-down reversal TEOM, BAM1020, OCEC
Negative Conc any size channel is negative GRIMM
Invalid Conc PM ≤ 0 or > 500 µg/m³ BAM1020
Invalid Carbon any carbon fraction ≤ −5 or > 100 µgC/m³ OCEC
Below MDL value at or below the method detection limit — advisory OCEC
Missing OC Thermal_OC or Optical_OC missing OCEC
Above MR concentration above the instrument's stated measurement range IGAC
Mass Closure Σ ions > PM2.5 IGAC
Missing Main any of NH₄⁺ / SO₄²⁻ / NO₃⁻ missing IGAC
Ion Balance cation/anion ratio outside 1.5 × IQR IGAC
Calibration Mode SAMPLE_TYPE != 1 Xact
Instrument Error ALARM in 100–110 Xact
Upscale Warning ALARM in 200–203 — advisory (the instrument's own warning class) Xact
Invalid Value element concentration outside 0–100 000 ng/m³ Xact
Internal Std Drift Nb outside ±20 % of its median Xact
High Uncertainty a normally-quantitative element reported above its detection limit with < 3σ confidence — advisory Xact
Negative any numeric column < 0 EPA

3. Aethalometers / BC monitors

AE33

Pattern [!ST|!CT|!FV]*[!log]_AE33*.dat
Native freq 1min
Parse read_table(delimiter=r'\s+', skiprows=5, usecols=range(67)); header is line 6, semicolon-suffixed names are stripped; index built from columns 0+1 (Date(yyyy/MM/dd) + Time(hh:mm:ss))
Quirks Files < 550 KB log "may not be a whole daily data". The pattern's [!ST|!CT|!FV] is a single-character negated class (excludes S T C F V | ! as the first character), not an alternation of ST/CT/FV prefixes — it happens to filter the ST/CT/FV log files but not for the reason it looks like.
Status column Status, mode bitwise. Raw file also carries ContStatus, DetectStatus, LedStatus, ValveStatus, which are not checked.
Error codes 1 tape advance / fast cal / warm-up · 2 first measurement (obtaining ATN0) · 3 stopped · 4 flow off by > 0.5 LPM · 16 calibrating LED · 32 calibration error · 1024 stability test · 2048 clean-air test · 4096 optical test
Deliberately not errors 128 / 256 (tape low warnings) and therefore 384 — data is still valid
QC rules Status Error, Invalid BC (0–20 000 ng/m³ over BC1–BC7), Insufficient; Invalid AAE added in _process
L2 output BC1BC7, abs_370abs_950, abs_550, AAE, eBC, BB(%) when present, QC_Flag

AE43

Identical to AE33 except:

  • Pattern [!ST|!CT|!FV]*[!log]_AE43*.dat; parsed with read_csv(parse_dates=['StartTime'], index_col='StartTime').
  • Keeps only the rows of the last SetupID in the file (a setup change mid-file discards the earlier segment).
  • ERROR_STATES is now identical to AE33's — it used to keep 384, which AE33 had dropped as a tape-low warning. A test pins the two lists equal.

BC1054

Pattern *.csv, native 1min
Parse header row located by scanning the first 20 lines for one starting with Time or Raw_Time (variants seen: header on line 1; Raw_Time,Time,…; a Data Report / User Report block plus 3 metadata lines; leading blank lines). Then read_csv(parse_dates=True, index_col=0, skiprows=skip); spaces stripped from column names; BC1 (ng/m3)BC1BC10, Flow(lpm)Flow, WS, WD, AT, RH, BP
Two clocks a file may carry both. Time is Met One's own field — "the date and timestamp for the data record. The timestamp is end of the minute" — written from the instrument's internal RTC, which an operator sets by hand (manual §3.5.7 SET CLOCK) and which can therefore be wrong. Raw_Time appears nowhere in the manual, so it comes from whatever logged or downloaded the file, on that host's clock.
Which is the index Raw_Time: the measurement happened at the wall-clock instant whatever the instrument believed, and in the corpus it is the well-behaved column — strictly increasing, no duplicates. The instrument clock is not: one fixture repeats timestamps and jumps 14 h 42 m mid-file, another sits a constant 12 h 04 m behind for all 1440 rows of a day (an RTC set to the wrong AM/PM, plus drift). The instrument's timestamp is kept as Instrument_Time, and a median disagreement over 60 s is warned — dropping it silently, as this reader used to, also discarded the only evidence that a clock needed resetting.
Quirks _QC first removes rows identical to their previous or next row (consecutive-duplicate filter) before flagging.
Status column Status, mode bitwise
Error codes 1 power failure · 2 digital sensor link failure · 4 tape move failure · 8 maintenance · 16 flow failure · 32 automatic tape advance · 64 detector failure · 256 sensor range · 512 nozzle move failure · 1024 SPI link failure · 2048 calibration audit · 65536 tape move
QC rules Status Error, Invalid BC (BC1–BC10), Insufficient, + Invalid AAE
L2 output abs_370abs_950, abs_550, AAE, eBC, plus every non-BC source column (Flow, AT, RH, BP, …), QC_Flag

MA350

Pattern *.csv, native 1min
Parse read_csv(parse_dates=['Date / time local'], index_col='Date / time local'); renames UV/Blue/Green/Red/IR BCcBC1BC5, Biomass BCcBB mass, Fossil fuel BCcFF mass, Delta-C, AAEAAE_ref, BB (%)BB
Status column Status, mode bitwise (rule is inert if the firmware labels it differently)
Error codes 1 power failure · 2 start up · 4 tape advance · 16 optical saturation · 32 sample timing error · 128 flow unstable · 256 pump drive limit · 2048 system busy · 8192 tape jam · 16384 tape at end · 32768 tape not ready · 65536 tape transport not ready · 262144 invalid date/time · 524288 tape error
QC rules Status Error, Invalid BC (BC1–BC5), Insufficient, + Invalid AAE
L2 output abs_375, abs_470, abs_528, abs_625, abs_880, abs_550, AAE, eBC, all non-BC columns, QC_Flag. Note the vendor's own AAE is preserved separately as AAE_ref.

4. Particle sizers

SMPS

Pattern *.txt, *.csv; native 6min
Parse opened with encoding='utf-8', errors='ignore' (real files contain non-UTF-8 bytes in the sample-path metadata). Header row is found by scanning for a first cell of Sample # (TXT / AIM 10.3, ~row 25) or Scan Number (CSV / AIM 11.x, ~row 52). Delimiter: tab for .txt, comma for .csv. Time from Date + Start Time, or DateTime Sample Start; date formats tried in order (%m/%d/%y, %m/%d/%Y, %Y/%m/%d for TXT; %d/%m/%Y for CSV). Some exports are transposed — if no date column is found the frame is rotated and re-parsed.
Size bins numeric column names → float; expected 11.8–593.5 nm. A mismatch warns; the file is only rejected when the caller passed an explicit size_range.
AIM reconciliation Two independent problems. (1) Mixed bin grids in one folder_partition_compatible_scans keeps the group with the most rows and names every dropped file in a warning. (2) Renamed metadataMETADATA_ALIASES rewrites AIM 11.x names to the AIM 10.3 canonical form (Total Concentration (#/cm³)Total Conc. (#/cm), Aerosol Temperature (C)Sample Temp (C), Aerosol Humidity (%)Relative Humidity (%), Aerosol Density (g/cm³)Density (g/cm), Impactor D50 (nm)D50 (nm), Test NameTitle, Geo. Std. DevGeo. Std. Dev., DMA Column transit time Tf (s)tf (s), DMA Exit to Optical Detector Td (s)td + 0.5 (s)). AIM 11.x-only columns keep their own names.
Status six columns, mode text, masks OR'd — whichever are present. The same information is reported under different names per host-software version:

Status Flag (10.3) ↔ Detector Status (11.x) — positive sentinel, OK when 'Normal Scan'
Instrument Errors (10.3) ↔ Classifier Errors (11.x) — error tokens, OK when empty
Communication Status (11.x only) — OK when '0'
Neutralizer Status (11.x only) — OK when 'ON'

Two value shapes: a positive sentinel column (OK when it equals a known-good string) and an error-token column (OK when empty, otherwise comma-separated fault names). 'Normal Scan' is auto-whitelisted on the token columns too, because some sites write the positive sentinel there instead of leaving it blank.
QC rules Status Error, Insufficient, Invalid Number Conc (2 000–1e7 #/cm³, NaN total ⇒ flagged), DMA Water Ingress (any bin ≥ 400 nm > 4 000 dN/dlogDp)
L2 output the dN/dlogDp matrix only (diameters in nm as columns) + QC_Flag. Statistics are not in the frame.
CPC the file names the counter that did the counting (Detector Model, Detector S/N, Nano Enhancer), reported in df.attrs as cpc_*. Its cut-off decides how far the lowest channels under-report — see Counting Efficiency. Not corrected for.
L3 sidecars {prefix}_dNdlogDp.csv, _dSdlogDp.csv (πd²·dN), _dVdlogDp.csv (πd³/6·dN), _stats.csv (from psd_stats). Pass append_stats=True to also append the statistics columns to the returned frame.
Was a gap an AIM 11.x export has neither AIM 10.3 name, so until all six were listed its Status Error rule could never fire — every scan passed the status check regardless of what the instrument reported. Verified on the CSV fixture: Detector Status='Normal Scan', Classifier Errors='Low aerosol flow', Communication Status='0', Neutralizer Status='ON' on all 62 scans, so the flag now fires (and ignored_status_errors=['Low aerosol flow'] clears it, as on the 10.3 dialect).

APS

Pattern *.txt; native 6min
Parse tab-separated; ~6 metadata lines (Sample File, Sample Time, Density, Stokes Correction, Lower/Upper Channel Bound) then a Sample # header row. Transposed exports are rotated (set_index('Sample #').T), and a failure there raises NotImplementedError with the original exception attached. Date from Date + Start Time, formats %m/%d/%y %H:%M:%S then %m/%d/%Y %H:%M:%S; the winning format is logged.
Sampling period the file header carries Sample Time (115 s in the corpus) — not a whole number of minutes, and not the 6min in the config. The native grid follows the detected period, so scans are not lost to bin collisions; df.attrs['raw_freq'] reports what was actually used.
Size bins numeric column names in 0.5–20 (µm) → float, rounded to 4 dp. Expected grid (0.542, 19.81, 51 bins); deviation warns loudly (an 8-year × 4-station audit of 1 485 files showed zero drift, so a deviation means a firmware change — and would reproduce the NaN-poisoned-concat problem SMPS had). The under-range <0.523 column is kept as metadata, not a bin.
Status column Status Flags, mode binary_string; OK is '0000 0000 0000 0000'
Bit meanings 0 laser fault · 1 total flow out of range · 2 sheath flow out of range · 3 excessive sample concentration · 4 accumulator clipped (> 65535) · 5 autocal failed · 6 internal temp < 10 °C · 7 internal temp > 40 °C · 8 detector voltage out of range · 9 reserved
QC rules Status Error, Insufficient, Invalid Number Conc (1–700 #/cm³)
L2 output / sidecars as SMPS, diameters in µm
Counting efficiency the APS under-counts at both ends — 85–99 % for solid particles, but falling from 75 % at 0.8 µm to 25 % at 10 µm for droplets (Volckens & Peters 2005). Not corrected for; see Counting Efficiency.

GRIMM

Pattern *.dat; native 6min
Parse read_csv(header=233, delimiter='\t', index_col=0, parse_dates=[0], encoding='ISO-8859-1', dayfirst=True) — the header row is a hard-coded line number; then columns 0–10 and the trailing 5 (or from column 128 for files named A407ST*) are dropped, and all values are divided by 0.035. Empty files are reported with print (not the logger) and skipped.
Status none
QC rules No Data (all channels NaN), Negative Conc (any channel < 0 — impossible for a counter), Insufficient
Deliberately absent concentration range limits, the equivalent of SMPS's MIN_TOTAL_CONC / APS's MAX_TOTAL_CONC. There is no GRIMM sample corpus to calibrate a plausible range against, and a wrong threshold silently deletes good data. The three rules above need no site-specific tuning.
Note like every reader, the completeness rule measures against the detected frequency (_resolved_freq), falling back to the config value only when detection fails.

5. Nephelometers, mass, chemistry, external sources

Aurora

Pattern *.csv; native 1min
Parse read_csv(low_memory=False, index_col=0); index coerced to datetime; column aliases 0°σspB/G/RB/G/R, 90°σspB/G/RBB/BG/BR, and Blue/Green/RedB/G/R, B_Blue/B_Green/B_RedBB/BG/BR; Raw_Data_Time dropped
Status first match among Status, status, Error, error, Flag, flag, S1 is renamed to Status; mode numeric, ok_value=0
QC rules Status Error, No Data, Invalid Scat Value (0–2 000 Mm⁻¹), Invalid Scat Rel (B < G & G < R), Insufficient
L2 output sca_550, SAE, plus all non-scattering columns (T1, T2, RH, P, S1, S2, …), QC_Flag
S1 is the status the production CSV (Data_Time, Raw_Data_Time, Red, Green, Blue, B_Red, B_Green, B_Blue, T1, T2, RH, P, S1, S2) contains none of the obvious status names — the status is S1, and it is now recognised. Evidence from 120 one-minute scans: S1 ∈ {0, 4}; the 17 rows with S1 == 4 are contiguous (00:06–00:22) and their scattering decays 185 → 0.78 Mm⁻¹ while S1 == 0 rows average 193 Mm⁻¹ — the signature of the zero/span check, where the instrument samples filtered air. Those rows pass the 0–2000 range check happily, so before this fix they were averaged into ambient means: on that fixture, dropping them raises the green-channel mean from 170.8 to 193.2 Mm⁻¹, i.e. a 13% low bias removed.
S2 moves in step (0x07 while ambient, 0xAB/0xA8 during the check) and is evidently a bitfield. Its individual bits are not decoded — no manual to hand, and guessing bit meanings would be worse than leaving it as data.

NEPH

Pattern *.dat; native 5min
Parse record-oriented, no header: read_csv(header=None, names=range(11)) then grouped by the record type in column 0. T = timestamp (YYYY MM DD HH MM SS across columns 1–6, zero-padded and concatenated). D = data; the NBXX sub-group is used, falling back to NTXX, columns 3–8 × 1e6 → B, G, R, BB, BG, BR. Y = state: col 2 pressure, 3 temp1, 4 temp2, 5 RH, 9 status. A file containing a record type outside {B,G,R,D,T,Y,Z} is skipped with a warning.
Quirks Y-row fields are attached to the data frame positionally (.values), so a file with unequal D and Y counts misaligns or raises.
Status column status (from Y row field 9, e.g. 0000), mode numeric, ok_value=0
QC rules / output same five rules and same outputs as Aurora

TEOM

Pattern *.csv; native 6min
Parse read_csv(skiprows=3, index_col=False); both alias maps are applied to every file because real exports mix conventions. remote/GUI names (Time Stamptime, System statusstatus, PM-2.5 base MCPM_NV, PM-2.5 MCPM_Total, PM-2.5 TEOM noisenoise, …) and usb/SNMP names (tmoStatusCondition_0status, tmoTEOMABaseMC_0PM_NV, tmoTEOMAMC_0PM_Total, tmoTEOMANoise_0noise, …) land on the same canonical short names. Format is then detected post-rename and logged: time present ⇒ remote (timestamps like 07 - 六月 - 2025 12:00:00, Chinese month names mapped, format='%d - %m - %Y %X'); Date+Time ⇒ usb. Neither ⇒ a file-named NotImplementedError.
Status column status, mode bitwise — this is the subtle one: the TEOM status condition register is a 32-bit bitfield (TEOM 1405 / 1405-F manual, Appendix A, Table A-1), so the instrument reports the decimal sum of every active warning. Treating it as a flat code (status != 0) makes any co-set bit unmaskable; bitwise mode lets one condition be whitelisted regardless of what else is set. ERROR_STATES = [1 << b for b in range(32)] — every bit counts as an error by default.
Documented bits 30 (1073741824) %RH high side A · 29 (536870912) dryer A · 28 (268435456) cooler A · 27 (134217728) exchange filter A · 26 (67108864) flow A · 25 (33554432) heaters side A · 24 (16777216) mass transducer A · 22–16 the same for side B · 14 user I/O · 13 FDMS device · 12 head 1 · 11 head 0 · 10 MFC 1 · 9 MFC 0 · 8 system bus · 7 vacuum pressure · 6 case/cap heater · 5 FDMS valve · 4 bypass flow · 3 ambient RH & temp sensor · 2 database log failure · 1 enclosure temp · 0 power failure
QC rules Status Error, High Noise (≥ 0.01), Non-positive, NV > Total, Spike, Insufficient
L2 output Volatile_Fraction = (PM_Total − PM_NV) / PM_Total plus every other columnOUTPUT_COLUMNS exists but is never applied

BAM1020

Pattern *.csv; native 1h
Parse read_csv(parse_dates=True, index_col=0, usecols=range(0, 21)); Conc (mg/m3)Conc; the literal value 1 is replaced with NA; then ×1000 (mg/m³ → µg/m³)
Status none
QC rules Invalid Conc (0–500 µg/m³), Spike
L2 output Conc, QC_Flag

OCEC

Pattern *LCRes.csv; native 1h
Parse read_csv(skiprows=3, on_bad_lines='skip'); time from Start Date/Time, formats %m/%d/%Y %I:%M:%S %p (RTCalc705 default) then %m/%d/%Y %H:%M:%S, then rounded to 1h. Three alias maps are applied unconditionally: RTCalc705 (Thermal/Optical OC (ugC/LCm^3)Thermal_OC, OC=TC-BC (ugC/LCm^3)Optical_OC, BC (ugC/LCm^3)Optical_EC, …), RTCalc802 (OC ugC/m^3 (Thermal/Optical)Thermal_OC, OptEC ugC/m^3Optical_EC, …), and the shared per-peak set (OCPk1-4-ug COC1_raw…, Pyrolized C ugPC_raw, ECPk1-5-ug CEC*_raw, Sample Volume Local Condition Actual m^3Sample_Volume). Firmware is inferred from the presence of per-peak columns and logged.
Derived at L1 OC{i} = OC{i}_raw / Sample_Volume (NaN on RTCalc705, which has no per-peak columns, or if Sample_Volume is missing — warned); PC = Thermal_OC − OC1 − OC2 − OC3 − OC4 when all four exist, else NaN
Status none
MDL Thermal_OC 0.3, Optical_OC 0.3, Thermal_EC 0.015, Optical_EC 0.015 µgC/m³ (class constant, not the config meta)
QC rules Invalid Carbon (−5 … 100 µgC/m³), Below MDL (value <= MDL), Spike, Missing OC
L2 output Thermal_OC, Thermal_EC, Optical_OC, Optical_EC, TC, OC1OC4, PC, QC_Flag

IGAC

Pattern *.csv; native 1h
Parse read_csv(parse_dates=True, index_col=0, na_values='-') with encoding='utf-8-sig'; column names stripped; everything coerced to numeric
Status none
MDL / MR from meta['IGAC'] — the vendor specification, 17 species covering gases and aerosol ions, exposed as reader.MDL / reader.MR. Species whose spec entry is None (HF, F⁻, PO₄³⁻ — not measured) are skipped everywhere. This replaced a second, disagreeing class-level copy of the 9 aerosol ions.
QC rules Mass Closureaerosol ions > PM2.5 when that column exists — gases are not part of the PM budget), Missing Main (NH₄⁺/SO₄²⁻/NO₃⁻), Above MR (any species above its stated measurement range), Ion Balance (cation/anion ratio outside 1.5 × IQR)
Below MDL a per-species diagnostic, not a flag: the log reports what fraction of each species sits below its limit, worst-first. (OCEC keeps a Below MDL flag, but at severity='warning' — same outcome by a different route. IGAC has 17 species, so a per-column count is more useful than one row-level flag.)
L2 output every spec species present in the file (gases included — they used to be dropped) + QC_Flag

Xact

Pattern *.csv; native 1h
Parse two header rows: line 0 is element names in caps (MAGNESIUM,,ALUMINIUM,,…), line 1 is the real header; the data rows carry one extra trailing field, absorbed as _extra_ and dropped. TIME parsed as %m/%d/%Y %H:%M:%S. Rows with Sample Type != 1 are dropped before rounding to 1h (the daily 00:00–00:30 QA check would otherwise displace a valid 00:30 sample). Element columns are matched by pattern (Mg 12 (ng/m3)Mg, Al Uncert (ng/m3)Al_uncert); environment columns are renamed to short forms (AT (C)AT, FLOW 25 (slpm)FLOW_25, …); PUMP START TIME, Output Pin 7, XC VER dropped.
Status column ALARM, matched by exact code, not bitwise; 0 = normal. decode_alarm() turns a code into text.
Alarm codes errors 100 X-ray voltage · 101 X-ray current · 102 tube temperature · 103 enclosure temperature · 104 tape · 105 pump · 106 filter wheel · 107 dynamic rod · 108 nozzle · 109 energy calibration · 110 software. Warnings 200 upscale Cr · 201 upscale Pb · 202 upscale Cd · 203 upscale Nb.
QC rules Calibration Mode, Instrument Error (100–110), Upscale Warning (200–203), Invalid Value (0–100 000 ng/m³), Internal Std Drift (Nb ±20 % of median). Each is registered only if its source column exists.
MDL the manual's own table (Operation Manual Appendix p.73, Xact.MANUAL_MDL) for the 29 elements it covers, selected by the file's SAMPLE_TIME — the limits vary ~8× between a 15-min and a 240-min sample, so one fixed number is right for only one configuration. meta['Xact']['MDL'] supplies the other 16, for which CES publishes nothing. Exposed as reader.MDL / reader.manual_mdl(minutes).
Uncertainty every element has a paired {element}_uncert column. The manual's limits are "interference free one sigma detection limits … at 68 % Confidence Level (C1σ) per US EPA IO 3.3 and Currie, 1968", and that uncertainty column is the same 1σ quantity — which is what lets Currie's criteria apply to it directly: detected at value ≥ 3σ, quantifiable at value ≥ 10σ.
Element reliability element_reliability(df) classifies each element from the fraction of samples meeting those criteria — quantitative / semi-quantitative / below-detection — and reports separately whether the manual publishes a limit for it at all (the two are independent: Nb, the internal standard, is measured superbly and has no published limit). Written to {prefix}_element_reliability.csv and summarised in the log. On the test corpus: 11 quantitative (S, K, Ca, Cl, Fe, Cu, Zn, Br, Mn, Pb, As), 7 semi-quantitative (Ti, Cr, Ni, Se, Ba, Bi, Sr), the rest below detection.
Below MDL a diagnostic, not a flag (same reasoning as IGAC, and more acute here): in the test fixture alone a dozen elements sit 100 % below their limit, so an "any element below MDL" rule would flag — and therefore NaN — every single row. The log reports the per-element fraction instead.
Why High Uncertainty is scoped the same trap. A rule firing when any element fails the 3σ test hits 96–100 % of rows on both fixtures, because a dozen elements are permanently below detection at any real site — a property of the element, not of the row. Scoped to elements this run measures well and the manual specifies, it fires on 0 % of clean rows and 8 % of the degraded fixture, where it caught S, K, Ca, Fe, Zn and Br degrading together: an instrument event rather than element noise.
L2 output all columns + QC_Flag

EPA — external, pre-aggregated

*.csv, native 1h, encoding='big5'. Accepts the Taiwan EPA hourly exports (測項 / 直式). Raises if more than one 測站 is present; drops 測站; pivots 測項/資料 into columns when present; renames AMB_TEMPAT, WIND_SPEEDWS, WIND_DIRECWD. Invalid-value markers are normalised by regex (…##, …L_) and then coerced to NaN. Columns reordered to SO2, NO, NOx, NO2, CO, O3, THC, NMHC, CH4, PM10, PM2.5, PM1, WS, WD, AT, RH. No status; one QC rule: Negative.

Removed: VOC and Minion

Both were pre-aggregated, second-hand data — somebody else's processed output, not an instrument's raw log — so there was no raw format for a reader to parse and the readers only ran generic checks. They were withdrawn from RawDataReader; RawDataReader('VOC', …) / ('Minion', …) now raises a KeyError carrying migration advice.

Read them yourself and go straight to the analysis functions:

import pandas as pd
from AeroViz import voc_potentials

df = pd.read_csv('voc.csv', index_col=0, parse_dates=True,
                 na_values=('-', 'N.D.'))
df.columns = df.columns.str.strip()
out = voc_potentials(df)      # validates species against support_voc.json

Minion arrives as a monthly report spreadsheet (read_excel), with a units row at the top and site-specific invalid markers (維護校正, Nodata, 0L); the handling depends on who produced that particular report, which is exactly why it does not belong behind a general-purpose reader.

Q-ACSM — reader not implemented yet

A real instrument, but no reader exists: no sample export has been available to write and test a parser against. It stays in meta (the 30-min native frequency is known) and is registered in supported_instruments.pending, so RawDataReader('Q-ACSM', …) raises a plain

NotImplementedError: Q-ACSM is a real instrument, but its reader is not
implemented yet — no sample export has been available to write and test a parser
against. Contribute one by adding AeroViz/rawDataReader/script/Q-ACSM.py with
_raw_reader and _QC ...

rather than the abstract-class TypeError it used to. Removing the pending entry is all that is needed once a reader lands.


6. Coverage gaps

Where the status/QC machinery was, or still is, doing less than it appears to:

Instrument Symptom Status
Aurora Status Error never fired — no obvious status column fixed: S1 recognised; the zero/span-check rows it marks were biasing ambient means 13% low
SMPS (AIM 11.x CSV) Status Error never fired — 10.3 names absent fixed: all six column names across both AIM dialects are checked
Any reader whose status column gets renamed degraded silently to "no errors ever" fixed: check_status_columns warns, naming what it looked for and what the file has
MA350 Status column presence never verified against a real export open — no fixture. The warning above will now say so at runtime instead of passing everything
GRIMM no concentration range check open by choice: no sample corpus to calibrate one against, and a guessed threshold silently deletes good data
Any reader using Insufficient partial first/last hours look sparse merely because the data starts mid-hour open — see P2-i in Data Levels §7

Full remediation list, with priorities and suggested order: Data Levels §7.


7. Summary matrix

Instrument Pattern Native Status column Mode QC rules _process
AE33 …_AE33*.dat 1 min Status bitwise 4
AE43 …_AE43*.dat 1 min Status bitwise 4
BC1054 *.csv 1 min Status bitwise 4
MA350 *.csv 1 min Status bitwise 4
Aurora *.csv 1 min Status (absent in practice) numeric 5
NEPH *.dat 5 min status (Y row f9) numeric 5
SMPS *.txt, *.csv 6 min Status Flag + Instrument Errors text ×2 4
APS *.txt 6 min Status Flags binary_string 3
GRIMM *.dat 6 min 3
TEOM *.csv 6 min status bitwise (32-bit) 6
BAM1020 *.csv 1 h 2
OCEC *LCRes.csv 1 h 4
IGAC *.csv 1 h 4
Xact *.csv 1 h ALARM exact code 5
EPA *.csv 1 h 1
Q-ACSM *.csv 30 min reader pending

Native is the meta['freq'] fallback; the grid actually used is detected per file at run time and reported as df.attrs['raw_freq'].

Three registry categories (config/supported_instruments.py):

  • supported — every row above except Q-ACSM: a module in script/ plus a meta entry.
  • pendingQ-ACSM: a real instrument, reader not written yet → NotImplementedError.
  • removedVOC, Minion: pre-aggregated second-hand data, no raw log to parse → KeyError with migration advice, see above.