Multi-wavelength Aethalometer (BC1054)
The BC1054 is a multi-wavelength aethalometer that measures black carbon concentrations at up to ten different wavelengths.
Data Format
- File format: CSV file
- Sampling frequency: Variable
- File naming pattern:
*.csv
- Data structure:
- Datetime index
- BC1-BC10 columns with units (ng/m³)
- Status column for instrument state
Measurement Parameters
The BC1054 provides black carbon measurements:
- BC1-BC10: Black carbon concentrations at different wavelengths (ng/m³)
Data Processing
Data Reading
- Processes CSV files with datetime index
- Standardizes column names
- Maps to consistent BC1-BC10 format
- Filters based on instrument status codes
- Handles duplicate timestamps
Quality Control
-
Instrument Status Check:
- Filters data based on instrument status codes:
- Status = 1: Power Failure
- Status = 2: Digital Sensor Link Failure
- Status = 4: Tape Move Failure
- Status = 8: Maintenance
- Status = 16: Flow Failure
- Status = 32: Automatic Tape Advance
- Status = 64: Detector Failure
- Status = 256: Sensor Range
- Status = 512: Nozzle Move Failure
- Status = 1024: SPI Link Failure
- Status = 2048: Calibration Audit
- Status = 65536: Tape Move
- Removes data points with other status codes indicating instrument problems
- Filters data based on instrument status codes:
-
Value Range Check:
- Removes physically impossible values:
- Negative concentrations
- Values > 20000 ng/m³
- Removes physically impossible values:
-
Data Representativeness:
- Requires at least 50% of expected data points in each 1-hour window
- Ensures temporal consistency of measurements
-
Complete Record Requirement:
- Ensures data completeness across all wavelengths
- Removes duplicate consecutive rows to prevent data redundancy
Output Data
The processed data contains:
- Time index: Data acquisition time
- BC1-BC10: Black carbon concentrations at each wavelength
- All measurements in ng/m³
Notes
- Provides real-time multi-wavelength measurements
- Enables detailed source apportionment
- Supports up to 10 wavelength channels
- Automatic tape advance system
- Status-based quality control